COMPANY X

Key machine specs

Reference section. These are the dealer's research-grade instruments, mostly high capex. They are kept for reference. The low-capex, high-value focus of this site is on the Floor Plan.

Figures gathered from the dealer posts, manufacturer pages and distributor listings. Each row links its source. Where sources disagree the note says so. Treat these as a shortlist aid and confirm exact figures and configuration on the dealer quote.

A Confirmed on a manufacturer page that was re-openedB Confirmed by a distributor, government page or two sourcesC Single secondary source or not re-openedD Unverifiable, conflicting, dead link or likely error

HORIBA LabRAM Odyssey

ParameterValueSourceTrustNotes
TypeFully automated confocal Raman microscope (2D and 3D imaging)dealer postCpage re-opened, model named; textual value not machine-matched
Spectrometer800 mm single-stagedealer postBpage re-opened (HTML); 100% of numeric tokens found; source type 'dealer post' is not the manufacturer
Wavelength range200-2200 nm achromatic spectrometer range (deep UV to NIR excitation lasers)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundHORIBA brochure (5 pages; printed 05/2026)
Low-frequency RamanDown to 5 cm-1; typical specification <10 cm-1 at 532/633/785 nmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundDepends on excitation wavelength
NanoRaman (TERS)Down to 10 nm resolution when coupled with a HORIBA AFMmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundOptional AFM coupling
LasersUp to 6 motorized, selectable 266-1064 nmdistributor brochureDdead/unreachable link (status 0)Laser set depends on configuration
Spectral resolution<=0.6 cm-1 (1800 gr/mm) and <=0.3 cm-1 (3000 gr/mm) at 532 nmdistributor brochureDdead/unreachable link (status 0)Depends on grating; not in the HORIBA brochure; HORIBA tech note RA-TN14 only explains that resolution depends on focal length (800 mm here) and grating groove density and gives no Odyssey figures; ask the dealer for the configured value
Sample stageAutomated stage up to 300 mm wafers (option)maker pageApage re-opened (HTML); 100% of numeric tokens foundOlder university units list a 75 x 50 mm XY stage
MappingSWIFT and DuoScan fast confocal mappingmaker pageCpage re-opened, model named; textual value not machine-matched
Size and weightAbout 1300 x 1194 x 473 mm, ~100 kgdistributor brochureDdead/unreachable link (status 0)Via search summary

Oxford Instruments Asylum Research Vero AFM

ParameterValueSourceTrustNotes
TypeAtomic force microscope using quadrature phase differential interferometry (QPDI)dealer postCpage re-opened, model named; textual value not machine-matchedDealer post names the maker as Oxford Instruments Asylum Research
Detection noise floorQPDI detector noise <10 fm/sqrt(Hz); QPDI lowers cantilever detection noise up to 10x or more vs optical beam deflectionmaker datasheetCpage re-opened (PDF text); 100% of numeric tokens found; row itself records conflicting values between sourcesMaker claim; brochure is 5 pages
Sensor noiseX&Y sensor <60 pm; Z sensor <50 pm; DC and AC height noise <15 pmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundAssumes required vibration and acoustic isolation
Sample sizeUp to 15 mm diameter and 7 mm thick (video-rate imaging typically ~3 mm)maker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Optical viewDiffraction limited (<1 um) NA 0.45; field of view 600 x 800 um; detector laser 650 nm spot <4 ummaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Model variantsVero S ambient; Vero ES sealed cell with environmental control; Vero VRS1250 video-rate (scan size shrinks from 30 um at <640 Hz to 500 nm at >1000 Hz)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundDealer only mentions Vero S and ES
ConfigurationsVero S (standard) and Vero ESdealer postCpage re-opened, model named; textual value not machine-matched
Scan range30 um X&Y; Z range >5 um (>2 um for video-rate scanning); out-of-plane motion <3 nmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundConfirms the earlier university-unit figure
SupportOne-year warranty; free technical supportdealer postCpage re-opened, model named; textual value not machine-matched

HORIBA Duetta

ParameterValueSourceTrustNotes
TypeSimultaneous fluorescence and absorbance spectrofluorometer (A-TEEM)dealer postCpage re-opened, model named; textual value not machine-matchedSome HORIBA pages call it 3-in-1
Fluorescence wavelength range250-1100 nm (CCD / spectrograph detector)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundHORIBA brochure (8 pages)
Absorbance wavelength range250-1000 nmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundNarrower than the fluorescence range
Light source75 W xenon arc lamp in a snap-in cartridge (LampSnap); wavelength accuracy +/-1 nmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Acquisition rateSpectral 510000 nm/min; EEM as fast as 1 s; A-TEEM as fast as 30 s (sample dependent)maker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Size and weight43 x 51 x 39 cm (W x D x H); 20.4 kgmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
PolarizersOptional at time of purchase (280-750 nm) on the Duetta-Bio version onlymaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundTable column reading is from extracted text; confirm which model has polarizers
Acquisition speedFull emission spectrum (EEM) in under 1 second with CCDdealer postCpage re-opened, model named; textual value not machine-matched
SensitivityWater Raman SNR >6000:1 RMS (350 nm excitation; 5 nm slits)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundMatches the dealer figure
Bandwidth optionsFluorescence (excitation and emission) 1/2/3/5/10/20 nm; absorbance 1/2/3/5/10/20 nmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundThird-party listing showed a shorter absorbance list (2/3/5/10 nm)
CorrectionReal-time primary and secondary inner filter effect correctiondealer postCpage re-opened, model named; textual value not machine-matched

Porometer POROLUX Revo

ParameterValueSourceTrustNotes
TypeGas-liquid capillary flow porometer using pressure-step/stability method (MP2 multistage pressure)dealer postCpage re-opened, model named; textual value not machine-matched
Pore size range13 nm to 500 ummaker pageApage re-opened (HTML); 100% of numeric tokens found
Max pressure35 bar (500 psi)distributor datasheetBpage re-opened (PDF text); 100% of numeric tokens found; source type 'distributor datasheet' is not the manufacturerConflict: one page shows 8 bar (likely another model); confirm on product sheet
Max flow200 L/mindistributor datasheetBpage re-opened (PDF text); 100% of numeric tokens found; source type 'distributor datasheet' is not the manufacturer
Suited materialsPolymeric membranes (flat sheet, hollow fibre), ceramic membranes, porous metals, nonwovensdealer postCpage re-opened, model named; textual value not machine-matched
Sample holder sizesNot foundmaker product sheetCvalue is a not-found placeholder; nothing to confirmSheet is behind a Porometer request form (labelled POROLUX 1000); older POROLUX 1000 and Cito use 13/25/47 mm holders but not confirmed for Revo; ask dealer for the technical specifications section

ModuleSci PicoEye-100 (PE-100)

ParameterValueSourceTrustNotes
TypeCompact tungsten-filament SEM with SE/BSE and optional EDSdealer postCpage re-opened, model named; textual value not machine-matchedModuleSci's own PE-100 pages (modulesci.com) call it both a tabletop and a floor-standing SEM; confirm the form factor on the quote
Resolution3.0 nm (tungsten filament source)maker pageCpage re-opened but only 100% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmedConfirmed by two ModuleSci pages; no PDF datasheet is published (contact@modulesci.com)
MagnificationUp to 300000x (effective over 100000x); dealer says 20x to 300000xmaker pageCpage re-opened but only 100% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmed
Accelerating voltage0.5-30 kV on one ModuleSci page; 1-30 kV on another and on the dealer postmaker pageCpage re-opened but only 66% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmedConflict between ModuleSci pages; confirm on quote
Max sample sizeDiameter 140 mm; height 45 mmmaker pageCpage re-opened but only 100% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmedNot in the dealer post
Detectors and optionsStandard ET-SE detector; options BSED (retractable BSE) EDS EBSD; chamber CCD and optical navigation camera; low-vacuum (LV) modemaker pageCpage re-opened but only 0% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmed
Sample exchangeWithin 60 secondsmaker pageCpage re-opened but only 100% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmed
Stage5-axis motorized: X 50 mm, Y 50 mm, Z 45 mm (dealer) or 50 mm (ModuleSci page), rotation 360 deg, tilt -20 to 45 degmaker pageCpage re-opened but only 100% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmedZ travel differs between sources
Vacuum systemRotary pump plus turbomolecular pumpdealer postCpage re-opened, model named; textual value not machine-matched
Dimensions680 x 780 x 1460 mm (W x L x H)dealer postBpage re-opened (HTML); 100% of numeric tokens found; source type 'dealer post' is not the manufacturer
CamerasNavigation camera and chamber cameradealer postCpage re-opened, model named; textual value not machine-matched

BioLogic BCS-900 series

ParameterValueSourceTrustNotes
ModulesBCS-905 (150 mA), BCS-910 (1.5 A), BCS-915 (15 A), BCS-975R (75 A) per channelmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundBrochure BT-BR-062-a (April 2025); the 975R is absent from the Sept 2024 brochure BT-BR-038-b
Channels per module8 independent channelsmaker pageCpage re-opened, model named; textual value not machine-matchedVia search summary
Total current (Sept 2024 brochure)Few uA to 120 A (up to +/-120 A with 8 BCS-915 channels in parallel)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundBrochure BT-BR-038-b; matches the dealer's 120 A figure
Total current (April 2025 brochure)15 uA to 300 A (up to 2 x +/-300 A with 4 BCS-975R channels in parallel)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundBrochure BT-BR-062-a; the two figures are different line-up versions not a contradiction; confirm modules in your quote
Current ranges per channelBCS-905: 5 ranges 100 mA to 10 uA; BCS-910: 5 ranges 1 A to 0.1 mA; BCS-915: 5 ranges 10 A to 1 mA; BCS-975R: 3 ranges 75 A to 750 mAmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundText extracted from PDF; verify the table in the PDF
Voltage0 to 10 V per cell (down to -5 V on /n variants); resolution down to 40 uV (18-bit) per dealermaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundVoltage range from maker brochure; 40 uV resolution is a dealer-post figure not found in the brochure text
Impedance (EIS)Native EIS 10 mHz to 10 kHz (BCS-905/910/915); 10 mHz to 2 kHz (BCS-975R)maker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Cabinet sizes6U / 12U / 24U / 38Udealer postBpage re-opened (HTML); 100% of numeric tokens found; source type 'dealer post' is not the manufacturer

Inovenso NE100

ParameterValueSourceTrustNotes
TypeSingle-nozzle electrospinning/spraying unit, bottom-up spinning, sealed safety cabinetmaker datasheetCpage re-opened, model named; textual value not machine-matchedInovenso Product Information sheet (2 pages, 2026)
Electrical0-40 kV (0.1 kV precision); max 0.125 mA; flow 0.35-7123 ml/h (0.01 ml/h precision)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundSame HV and pump block on NE100/200/300/NS24
Collectors and distanceStationary plate 370 x 130 mm; nozzle-collector distance 11-211 mm; drum collector up to 500 RPM (per text)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundMaker text mentions the drum; spec table lists only the plate
Interface and size9-inch touch screen, Wi-Fi remote monitoring, recipe save/recall, CE; 780 x 680 x 1050 mm; ~150 kgmaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundThird-party catalogues had a different screen size

Inovenso NE200

ParameterValueSourceTrustNotes
TypeSingle-nozzle lab unit with flat collector and rotating drum; budget modelmaker datasheetCpage re-opened, model named; textual value not machine-matchedInovenso Product Information sheet (2 pages, 2026)
Electrical0-40 kV; max 0.125 mA; flow 0.35-7123 ml/hmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Drum collector100 mm x 150 mm drum (314 x 150 mm deposition); 100-500 RPM; aluminiummaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Distance and sizeNozzle-collector 33-233 mm; plate 370 x 130 mm; 780 x 680 x 1050 mm; ~150 kg; optional atmosphere controlmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found

Inovenso NE300

ParameterValueSourceTrustNotes
TypeMulti-nozzle lab unit: up to 8 nozzles in 2 feeding sets of 4, also single-nozzle; coating homogeneity systemmaker datasheetCpage re-opened, model named; textual value not machine-matchedMaker sheet says 8 nozzles; one third-party listing said 9
Electrical0-40 kV; max 0.125 mA; flow 0.35-7123 ml/hmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Drum collector100 mm x 220 mm drum (314 x 220 mm deposition); 100-500 RPM; X-axis traverse 2-20 mm/s over 30-80 mmmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
Distance and sizeNozzle-collector 31-231 mm; min solution 1 ml per nozzle, 16 ml full load; 780 x 680 x 1050 mm; ~150 kgmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found

Inovenso NS24 (Nanospinner24)

ParameterValueSourceTrustNotes
TypeMost versatile lab system: up to 12 nozzles in 2 sets of 6, also single-nozzle; 50+ units installed per makermaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundInovenso Product Information sheet (2 pages, 2026)
Drum collector120 mm x 280 mm drum (376.8 x 280 mm deposition); 100-2000 RPM; X-axis traversemaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundFaster drum than NE200/NE300
Electrical and size0-40 kV; max 0.125 mA; flow 0.35-7123 ml/h; nozzle-collector 21-221 mm; 780 x 680 x 1050 mm; ~150 kgmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found

Inovenso NS1

ParameterValueSourceTrustNotes
TypeEntry-level single-nozzle unit, open spinning area (fits in a fume hood), 4.3-inch touch screenmaker datasheetCpage re-opened but only 0% of value tokens found in extracted text (JS-rendered, image or different formatting possible); not confirmedInovenso Product Information sheet (2 pages, 2026)
Electrical and size0-30 kV; max 0.170 mA; flow 0.35-7123 ml/h; plate 300 x 238.5 mm; distance 30-320 mm manual; 760 x 645 x 440 mm; ~40 kgmaker datasheetApage re-opened (PDF text); 91% of numeric tokens found

Inovenso NS Starter Kit

ParameterValueSourceTrustNotes
TypeLow-cost single-nozzle kit; 0-30 kV, 0.170 mA; syringe pump 10 nl/min to 120 ml/min; ~2.2 kgmaker datasheetApage re-opened (PDF text); 80% of numeric tokens foundNot in the dealer's featured list; shown for completeness

AID GmbH multiSpot

ParameterValueSourceTrustNotes
TypeCombined EliSpot/FluoroSpot reader and automated microscope (model MSR08)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundAID Product Overview 2025 (Rev.30, 17 pages), multiSpot = model MSR08, page 12
Objectives4x, 10x, 20x on a software-controlled changer (others on request)maker datasheetApage re-opened (PDF text); 100% of numeric tokens foundOnly reader in the AID range with objectives
Plate formats96- and 384-well plates, up to 4 glass slides, Terasaki plates; 1 plate per runmaker datasheetApage re-opened (PDF text); 100% of numeric tokens found
CameraColour digital FireWire, 5 and 2 MP, fluorescence-optimisedmaker datasheetCpage re-opened, model named; textual value not machine-matchedMatches the dealer post
Illumination and filtersLED ring, two XBO sources, 3&1 filter wheel; FITC, Cy3, Cy5 narrow-band filters on board plus quad band; 1-3 colour analysismaker datasheetCpage re-opened, model named; textual value not machine-matchedFluoroSpot max 3 colour overlays
Read timeAbout 2 min (EliSpot, 96-well); under 7 min (2-colour FluoroSpot, 96-well)maker datasheetApage re-opened (PDF text); 100% of numeric tokens found
ApplicationsEliSpot, FluoroSpot, cell counting/viability, HEp-2 screening, HLA screening; no colony countingmaker datasheetCpage re-opened, model named; textual value not machine-matched
Power and approvalsMax 750 mA at 24 V DC; CE marked; high-end PC with QHD 27-inch monitormaker datasheetApage re-opened (PDF text); 100% of numeric tokens foundMatches the dealer post